H-1200 Fast Fluid Warmer with integrated Air Detector / Clamp and H-31B and H-30 Air Detector / Clamp Accessory

Notice type: Warning

Date: 31/01/2008

 

Product name or type:
H-1200 Fast Fluid Warmer with integrated Air Detector / Clamp and H-31B and H-30 Air Detector / Clamp Accessory


Reference:
SN2008(02)


Manufacturer Or Supplier:
Smiths Medical, United Kingdom


Target Audience:
Accident & Emergency Departments; Intensive Care Units; Cardio Thoracic Units; All Wards; Clinical Perfusionists; IV Nurse Specialists; Maternity Units; Theatre Managers; Theatre Nurses; Risk Managers; Clinical Engineering


Problem Or Issue:
The risk of air embolism if the power of the fluid warmer unit is interrupted either manually or through power failure during an air detection event.


Background Information Or Related Documents:
Smiths Medical circulated an advisory notice to Irish hospitals in July 2007 advising of the potential risk of air embolism with the above device.  They advised that this could occur when the power supply to the unit is interrupted during an air detection alarm e.g. by manually switching off the power during an air detection event, causing the air detector clamp to open and remain open.  Smiths Medical provided Irish hospitals with a ‘quick reference guide’ warning them of this issue.   

Despite this action, an incident was reported in the USA, where during an air detection event the alarms were activated but the user turned the unit off despite the recommendations of the ‘quick reference guide’.  Air was delivered to the patient but no serious patient injury was reported.

H-1200 Fast Fluid Warmer with integrated Air Detector / Clamp and H-31B and H-30 Air Detector / Clamp Accessory Document


Actions To Be Taken:

Further to this incident the Irish Medicines Board (IMB) advises users to discontinue the use of the device until further notice.  

  • Identify and quarantine affected devices until further notice.  
  • If no alternative fluid warmer is available, ensure users are aware of the advice provided in the manufacturer’s customer letter (copy attached as appendix 1) and ensure that the quick reference guide is attached to all affected devices.  
  • Contact the manufacturer to request a copy of quick reference guide if not attached to all affected devices


Further Information:
All adverse incidents relating to a medical device should be reported to the:   

Health Products Regulatory Authority
Medical Devices Department 
Kevin O’Malley House 
Earlsfort Centre 
Earlsfort Terrace 
Dublin 2   

If you have any enquiries, you may contact the Medical Devices Department at:   

Telephone:        +353-1-6764971 
Fax:                  +353-1-6344033 
Email:               vigilance@imb.ie 
Website:            www.hpra.ie


SN2008(02): HPRA Safety Notice - H-1200 Fast Fluid Warmer with integrated Air Detector / Clamp and H-31B and H-30 Air Detector / Clamp Accessory

SN2008(02): Appendix 1 - Smiths Medical Safety Alert


« Back