K- Series, Competition, Compact, Compact – Hemi Combi Drive, Kid and Ultra Light Wheelchairs

Notice type: Warning

Date: 20/11/2006

 

Product name or type:
K- Series, Competition, Compact, Compact – Hemi Combi Drive, Kid and Ultra Light Wheelchairs


Reference:
SN2006(08)


Manufacturer Or Supplier:
Invacare


Target Audience:
Occupational Therapists Physiotherapists; Public Health Nurses; Risk Managers; Nursing Home Managers


Problem Or Issue:
The wheelchair seat covers manufactured by Invacare between February and October 2003 were supplied with a covering material, which did not meet the requirements for flammability defined by the manufacturer.


Background Information Or Related Documents:

In November 2003, a recall of the above product was initiated by Invacare following the discovery that seat covers manufactured, for the following list of devices, between February 2003 and October 2003 did not comply with the flammability requirements.   

  • The K-series  
  • Competition  
  • Compact  
  • Compact – hemi drive 
  • Kid 
  • Ultra Light 
Invacare have had no reported incidents relating to this flammability issue.  The manufacturer has confirmed that when the chair is used with a cushion, the risk is reduced.   

The affected product was supplied to the Irish market by Lifestyle Mobility and Irish Wheelchair Association.  Since this time the supply of  Invacare products has changed from the Irish Wheelchair Association to Invacare Ireland.   

The IMB has not been able to obtain confirmation from the manufacturer indicating that all affected product in Ireland have been checked and new covers provided and fitted.

K- Series, Competition, Compact, Compact – Hemi Combi Drive, Kid and Ultra Light Wheelchairs Document


Actions To Be Taken:
Ensure that the purchaser and prescribers of these devices within your healthcare institution:  

1. Identify all K-series, competition, compact, compact-hemi drive, kid and Ultra light seat covers that were supplied to your healthcare institution in 2003 

2. Arrange with your supplier (Invacare Ireland or Lifestyle Mobility) to check the seat cover material.  The inside of the non-flame retardant cover is black while the inside of the flame retardant cover is grey. 

3. Ensure that all affected seat covers are replaced with new covers.


Further Information:
All adverse incidents relating to a medical device should be reported to the:  

Health Products Regulatory Authority
Medical Devices Department 
Kevin O'Malley House 
Earlsfort Centre 
Earlsfort Terrace 
Dublin 2   

If you have any enquiries, you may contact the Medical Devices Department at:   

Telephone:     +353-1-6764971 
Fax:               +353-1-6344033 
Email:            medicaldevices@imb.ie
Website:        www.hpra.ie 

Enquiries to the manufacturer should be addressed to:   

Declan Services 
Invacare Ireland  

Telephone:     +353-87-6894940 
Email:            dservices@invacare.com

SN2006(08): K- Series, Competition, Compact, Compact – Hemi Combi Drive, Kid and Ultra Light Wheelchairs


« Back