Zoll AED Plus Defibrillator
Notice type:
Advisory
Date:
23/05/2011
Product name or type:
Zoll AED Plus Defibrillator
Reference:
IMB Safety Notice: SN2009(13) Updated Version
Manufacturer Or Supplier:
Zoll Medical Corporation
Target Audience:
General Practitioners
General Public
Community First Responder schemes
Risk Managers
Hospitals
Nursing Homes
Schools
Sports clubs
Problem Or Issue:
The Zoll AED Plus Defibrillator may not deliver therapy
Background Information Or Related Documents:
Zoll received reports that following a long period without use (typically greater than four years); a ZOLL® AED Plus® may prompt “change batteries” during use and fail to deliver therapy. Turning the device off completely, waiting ten seconds for the unit to re-set and then turning it back on is necessary to resume proper operation.
Zoll’s investigation of this problem has determined that some batteries may develop high internal resistance that interferes with the batteries’ expected performance, lengthening charging time beyond specified and clinically acceptable limits.
In February 2009, ZOLL Medical Corporation started to inform owners of all AED Plus devices with serial numbers below X_ _ _200000 to replace batteries that have been installed in devices for a period longer than three (3) years and at subsequent three (3) year intervals.
Units above serial number X_ _ _200000 are not affected by this corrective action as they contain software that can detect this battery condition and identify when batteries require replacement. In lieu of adding the battery replacement label, users may obtain this new software for their AEDs at no cost at
www.ZOLLAEDPlusbatteryhelp.com
Four distributors have supplied the affected product in Ireland, Pulse Medical Ltd, Heartbeat Safety Ltd, AED Direct and Medilink Services. Zoll has been working with the distributors to ensure that the issue is communicated to all users of the product in Ireland. Zoll have been unable to confirm to the IMB that the recommended actions have been completed on all devices.
Zoll AED Plus Defibrillator Document
Actions To Be Taken:
1. Ensure that the relevant personnel in your organisation are made aware of this
issue.
2. Determine if you have purchased the affected serial numbers of this product,
AED Plus devices with serial numbers below X_ _ _200000.
3. If you have an affected device confirm that the recommended actions have been completed:
Ensure that the date when the batteries were installed in your device has been checked.
a. If batteries were installed more than three (3) years ago, replace the batteries
as soon as possible and add the label provided by Zoll showing the next
scheduled replacement date. Read and add the addendum incorporating this
information to the administrator’s guide that was provided by Zoll.
b. If batteries are not more than three (3) years old, add the label provided by
Zoll completed with the next scheduled replacement date. Read and add the
addendum, provided by Zoll, incorporating this information to the
administrator’s guide.
Further Information:
All adverse incidents relating to a medical device should be reported to the:
Health Products Regulatory Authority
Kevin O’Malley House
Earlsfort Centre
Earlsfort Terrace
Dublin 2
Telephone: +353-1-6764971
Fax: +353-1-6344033
E-mail:
vigilance@imb.ie
Website:
www.hpra.ie
Enquiries to the manufacturer should be addressed to:
Manufacturer:
Zoll Medical Corporation
269 Mill Road
Chelmsford
Massachusetts
USA
Telephone: + 31 651937017
E-mail:
erozeboom@zoll.com
Distributor:
Pulse Medical Ltd
Unit 8
Ballyboggin Business Centre
Ballyboggan Road
Glasnevin
Dublin 11
Ireland
Telephone: +353 1 8602631
Fax: +353 1 8602640
Heartbeat Safety Ltd.
Unit 3
Crookstown Business Park
Crookstown
Co Kildare
Ireland
Telephone: +353 59 8623975
Fax: +353 59 8623870
Medilink Services
Link House
81 Sydenham Road
Belfast
Northern Ireland
Telephone: +44 2890 582999
Fax: +44 2890 582900
Please click here to download a PDF version of the Safety Notice
« Back
Date Printed: 28/03/2024