pH Indicator Strips (non-bleeding) pH 0-6.0 & 2.0-9.0 Priority 2 – Warning Caution In Use
Notice type:
Advisory
Date:
09/05/2014
Product name or type:
pH Indicator Strips (non-bleeding) pH 0-6.0 & 2.0-9.0
Reference:
SN2014(24)
Manufacturer Or Supplier:
Merck Serono
Target Audience:
All Hospital Staff
All Nursing Home Staff
A&E Departments
Carers
Chief Executive Officers
Clinical Directors
Day Surgery Units
Emergency Medical Technicians
Gastroenterology Departments
General Practitioners
General Public
Healthcare professionals who use these devices
Healthcare professionals managing patients who use these devices
High Dependency Units
Hospital Managers
Hospital Pharmacists
Intensive Care Units
Maternity Units
Midwifery Departments
Neonatology Departments
Nursing Managers
Nursing staff
Obstetrics and Gynaecology Departments
Oncology Nurse Specialists
Paediatric Departments
Paramedics
Purchasing / Procurement / Material Managers
Risk Managers
Supplies Managers
Problem Or Issue:
A communication issued by Merck Serono to Irish customers in May 2013 indicated that an additional ‘safety feature’ had been introduced for these products, to facilitate their use in non-buffered solutions.
Merck Serono has informed the Irish Medicines Board (IMB) that the safety feature introduced in 2013 does not function as intended over the lifetime of the product.
The manufacturer has issued the attached Field Safety Notice (FSN) to users of these devices to reiterate that these devices should only be used with gastric aspirate. Merck Serono is also implementing a revision to the Instructions For Use (IFU) for their pH Indicator Strips in order to strengthen the warning regarding use with low buffered solutions.
The IMB wishes to remind users to use these devices with caution.
Background Information Or Related Documents:
pH Indicator Strips (non-bleeding) pH 0-6.0 & 2.0-9.0 Document
Concerns have been expressed to the IMB in relation to the performance of Merck Serono pH Indicator Strips (non-bleeding). The IMB wishes to remind users to use these devices with caution. These strips are designed to verify safe nasogastric tube placement by measuring the pH of human gastric aspirate using a colour change reaction. Patient death can occur if a patient is fed using a tube that has been incorrectly positioned in the lung.
The IFU indicates that these devices are only for use with gastric aspirate.
Actions To Be Taken:
The IMB advises users:
(1) Merck Serono pH Indicator strips (0-6.0 product code 090120604 and 2.0-9.0 product code 090120603) should be used with caution.
a. These strips are only for use with gastric aspirate
b. These strips are not suitable for use with low buffered solutions
c. These strips are not suitable for use with any kind of water (distilled, sterilised, potable, bottled or tap water)
(2) Forward this IMB Safety Notice to all those within your organisation that need to be aware of this information. Please also pass this Safety Notice on to any end users or organisations where these devices may have been distributed.
Further Information:
Enquiries should be addressed to:
Merck Serono Ltd
Bedfont Cross
Stanwell Road
Feltham
Middlesex
TW14 8NX
United Kingdom
Telephone: +44 (0) 208 818 7373
Fax: +44 (0) 208 818 7263
E-mail:
medinfo.uk@merckgroup.com
Website:
www.merckserono.co.uk
All adverse incidents relating to a medical device should be reported to:
Health Products Regulatory Authority
Human Products Monitoring
Kevin O’Malley House
Earlsfort Centre
Earlsfort Terrace
Dublin 2
Telephone: +353-1-6764971
Fax: +353-1-6344033
E-mail:
devicesafety@hpra.ie
Website:
www.hpra.ie
Please click here to view a copy of the field safety notice