Actions To Be Taken:
Steriliser Selection
Users of sterile medical devices have three initial options to consider:
- employ sterile single use devices,
- have reusable devices sterilised by a certified Sterile Services Department (SSD) or
- reprocess the devices themselves.
If reprocessing reusable devices is the option chosen, the choice of bench-top steriliser should be based on the suitability of the steriliser for the type of devices to be sterilised e.g. wrapped / unwrapped. A steam steriliser unit should be chosen with adequate capacity to sterilise the required number of devices at the desired frequency. Advice should be sought from the steriliser manufacturer or an independent advisor, who should be able to confirm whether a particular bench-top steriliser is suitable for sterilising particular instruments and meets the user requirements. The types of bench-top steam sterilisers are:
Vacuum sterilisers (See Table 1)
- B-cycle sterilisers are capable of effective sterilisation of pre-wrapped devices, porous devices such as fabrics, and devices that are hollow or have lumen.
- S-cycle sterilisers only process devices specifically as advised in the manufacturer’s instructions.
- Vacuum sterilisers have a vacuum pump that causes forced air removal from the chamber and assists steam penetration, but they require a post-sterilisation drying phase which increases cycle duration.
- The most common type of steriliser used is a B-cycle steriliser.
Non-vacuum, bench-top sterilisers (See Table 1)
- Non-vacuum, bench-top sterilisers utilise an N-cycle.
- They are NOT capable of sterilising porous devices, devices that are hollow, have lumen or are pre-wrapped. This is because in non-vacuum sterilisers, the chamber air is passively displaced by the steam entering the chamber and the displacement of air from ‘hard to access’ devices can be ineffective, causing reduced steam-instrument contact.
- They should only be used for solid, unwrapped instruments for immediate use only, where the devices are transported aseptically to point of use. This is because devices risk becoming contaminated once the steriliser is opened.
Table 1: Bench-topSteriliserCycle Types | SteriliserType: | Non-Vacuum | Vacuum | Cycle Type: | N | B | S | Air Removed by: | Passive displacement (using steam) | Vacuum (forced air removal) | Vacuum-pulsed (forced air removal) | Devices that can be sterilised: | Solid only | Solid Hollow | Solid Hollow | Unwrapped only | Wrapped Unwrapped Porous | Wrapped Unwrapped | Notes | Most basic & least expensive. Totally unsuitable for wrapped devices or devices in pouches. Devices must be used immediately. | More complex & expensive. Most common. | Expensive to purchase and maintain. Specialiststerilisersto sterilise devices specified by the manufacturer. | |
Installation and Validation
Whilst bench-top sterilisers undergo some degree of validation and testing prior to leaving the manufacturing facility, purchasers of sterilisation units should check what installation, commissioning and validation is required before the steriliser is put into service.
- The bench-top steam steriliser should be installed and commissioned, by testing to the relevant standard, by the manufacturer or suitably qualified field service technician or engineer before first use.
- Pressure testing should be completed by a suitably qualified person, approved by the organisation’s insurance provider.
- Certification should be obtained that the steam steriliser is fit for use.
- Manuals on the operation and maintenance of the steriliser should be provided at installation and available at the point of use.
- On-the-job staff training should be performed at installation by a suitably qualified field service technician and documented.
- A logbook should be initiated and maintained to document the person responsible for and the date of: installation, pressure vessel testing, sterilisation batches, cleaning, maintenance and servicing etc.
- Details of sterilisation batches such as cycle parameters and devices sterilised should be documented in a sterilisation record,
- Procedures or SOPs should be written that capture all quality system elements including operation, training, verification, failures, cleaning, record keeping, product release, calibration, maintenance, servicing, auditing etc., identifying the person(s) responsible for each.
- All documentation/records should be retained for review and audit.
Loading
The bench-top steriliser must be loaded according to the manufacturer’s instructions and relevant code of practice.
- Items to be sterilised must be pre-cleaned e.g. surfaces cleaned and dried, preferably in an automated washer/disinfector for safety reasons. If the instrument is not clean, it cannot be sterilised.
- Instructions for use should be available at the point of use.
- Chamber trays should not be overloaded beyond the recommended loading capacity, as this will adversely affect steam circulation causing ‘cold-spots’ that reduce the sterilisation efficiency.
- Bladed devices such as scissors should be placed on chamber trays in an open position and bowls should only be placed either upside down or tilted downwards.
- Instruments must not be stacked but must be loaded to expose the greatest surface area possible to the steam.
- The chamber door should be securely closed, to prevent serious injury due to sudden changes in the chamber pressure.
Sterilisation Cycle
- The sterilisation cycles should be validated for use with particular RIMDs, as per the manufacturer’s recommendations, and should define the critical cycle parameters such as sterilisation temperature, duration.
- The air removal effectiveness of vacuum steam sterilisers should be verified daily using a suitable test e.g. the Bowie-Dick, Helix Test etc.
- All devices must be allowed to dry and cool down before they are removed from the steriliser, to prevent microbial growth.
- Sterilising cycle records should be kept for traceability purposes.
Storage
- If the medical device must be sterile for use, it may be:
- wrapped prior to sterilisation and stored in its wrapping to maintain sterility, in a location where its sterility will not be compromised or
- stored clean until required and sterilised immediately prior to use.
- Ensure that clean and dirty instruments are kept in distinctly segregated areas to prevent cross-contamination or mix-ups.
Water Quality
- Water that contains contaminants can have a serious effect on the patient. Contaminants present on sterile instruments can be directly introduced into parts of the body that are normally protected by skin or mucous membranes.
- Contaminants include harmful micro-biological substances that are resistant to sterilisation (e.g. endotoxins) and water impurities (e.g. metal deposits).
- Water that contains impurities can adversely affect the quality of the steam used in the sterilisation cycle causing:
- staining and pitting corrosion of the instruments and steriliser
- build-up of residue on the steriliser’s elements, reducing heat transfer.
- The water reservoir should be:
- emptied daily after use, rinsed (ideally with sterile water) and left to dry overnight to avoid standing water.
- Refer to the manufacturer’s recommendations to determine the minimum acceptable water quality, to ensure low levels of micro-organisms and impurities.
Maintenance
- Routine maintenance should be carried out, as per the manufacturer’s instructions, to monitor and ensure consistent steriliser performance.
- Maintenance procedures should be documented and available at the point of use.
- Steriliser cleaning should be carried out as per the manufacturer’s recommendations. The purpose of surface cleaning is to minimise the build-up of:
- dust particles on the exterior steriliser surface that may result in cross-contamination
- residue on the interior steriliser surface that may reduce sterilisation efficiency and effectiveness.
- Maintenance of water quality is as listed in the Water Quality section.
- Completed maintenance operations should be recorded in a logbook.
Servicing
Bench-top sterilisers must be maintained and serviced according to the manufacturer’s recommendations.
- Bench-top sterilisers should be fully serviced by a suitably qualified service technician, in accordance with the manufacturer’s instructions, industry standards and relevant codes of practice. Note that the frequency of servicing (e.g. yearly, quarterly etc.) will also depend on the manufacturer’s instructions and industry standards.
- The service should include (i) equipment cleaning, (ii) filter, equipment, control valve maintenance or replacement and (iii) equipment operation & calibration.
- All servicing activities and the results should be documented.
In following recommended practices, standards and legislation that cover steriliser installation, validation, operation, maintenance, training, safety and insurance, the risks to users, owners, operators and patients are significantly reduced.