Follow @TheHPRA
Contact us
As Gaeilge
My HPRA:
Login
Register
About Us
Medicines
Veterinary
Medical Devices
Cosmetics
Controlled Substances
Blood, Tissues, Organs
Medical Devices
Safety Information
Safety communications
My HPRA:
Login
Register
About Us
Medicines
Veterinary
Medical Devices
COVID-19 Updates
Our role
Safety Information
Reporting safety issues
Monitoring device safety
Safety communications
About HPRA Safety Notices
Manufacturer's Field Safety Notices
User Engagement
Consumer advice
Regulatory Information
Registration
Medical Devices News & Events
Special Topics
Contact Medical Devices
Cosmetics
Controlled Substances
Blood, Tissues, Organs
Silicone Gel filled Breast Implants manufactured by Poly Implant Prothese (PIP). (All models and lot numbers) 05.10.2010
Notice type:
Advisory
Date:
05/10/2010
Product name or type:
Silicone Gel filled Breast Implants manufactured by Poly Implant Prothese (PIP). (All models and lot numbers).
Reference:
IMB Safety Notice: SN2010(13)
Manufacturer Or Supplier:
Poly Implant Prothese (PIP)
Target Audience:
Medical directors
Plastic surgeons and all surgeons involved in breast reconstruction
Directors of surgical units involved in breast construction
General practitioners
Practice managers
Practice nurses
Private Cosmetic Clinics
Problem Or Issue:
A recall of silicone gel filled breast implants manufactured by the French company Poly Implant Prothese (PIP) was initiated in Ireland on the 30th March 2010. The recall follows a manufacturing site inspection in France which identified unauthorised silicone gel being used in the product.
Background Information Or Related Documents:
In March of this year the IMB was informed by the French medical device regulatory authority (AFSSAPS) that they suspended Poly Implant Prothese (PIP)’s the marketing, distribution and export of these products. This action was taken following an inspection of PIP’s manufacturing plant where it was found that most implants manufactured since 2001 were filled with an unauthorised silicone gel which differs in composition from the originally approved material.
The inspection conducted by AFSSAPS was prompted when an increase in the number of incident reports regarding device rupture and local complications associated with these products was noted in France.
At the time of the recall the IMB ensured that all Hospitals /Clinics in Ireland that used these devices were advised of the issue. The IMB advised these facilities to cease the use of these products.
To date the IMB has received one confirmed adverse incident associated with the use of this device.
Since the issue was identified in March the IMB has closely monitored the situation in liaison with our colleagues in Europe to determine potential safety implications.
Tests on the affected implants have been conducted by several Regulatory Agencies, including the Therapeutic Goods Administration (TGA) in Australia, Medicines and Healthcare products Regulatory Agency (MHRA) in the UK and French Regulatory Agency, Afssaps.
On July 2010 the Department of Health and Ageing, Therapeutic Goods Administration in Australia issued the following results
https://search.tga.gov.au/s/search.html?collection=tga-websites-web&query=alerts%20devices%20silicone
The results of TGA testing to date indicate that the PIP breast implants supplied in Australia conform to the relevant international standards for this type of product including those for gel cytotoxicity and shell strength.
On the 3rd September 2010 the Medicines And Healthcare products Regulatory Agency in the UK advised the following
http://www.mhra.gov.uk/NewsCentre/Pressreleases/CON093706
The tests found no evidence of genotoxicity (potential for cancer) or chemical toxicity of the filler material in the implants.
On the 4th October 2010 the Medicines And Healthcare products Regulatory Agency in the UK issued the following recommendations to Surgeons
http://www.mhra.gov.uk/Publications/Safetywarnings/MedicalDeviceAlerts/CON096755
On the 27th September the French Regulatory Agency AFSSAPs released their
investigation findings.
http://ansm.sante.fr/
The physicochemical analysis confirm that gels filling tested breast implants of PIP company are not those described in the manufacturer's design file.
Two tests of mechanical strength are compliant with requirements of existing standards for breast implants. However the test for elongation until rupture is not in accordance with the standards.
A test shows that the gel of PIP breast implants has no acute toxic effects on tissues (cytotoxicity).
The results of the intradermal irritation test show an irritant behaviour of PIP gel that is not found on other silicone gels of other breast implants.
Three tests for assessing possible effect of the PIP gel on DNA of the cells (genotoxicity) were carried out, 2 in vitro and 1 in vivo in mice. If both in vitro tests have shown negative results, finding obtained in vivo do not allow in the present state to conclude on the absence or presence of a genotoxic effect.
Further testing will be conducted by Afssaps with results expected early in 2011. The IMB will continue to closely monitor this issue with AFSSAPS and our other regulatory colleagues
Silicone Gel filled Breast Implants manufactured by Poly Implant Prothese (PIP). (All models and lot numbers) 05.10.2010 Document
Actions To Be Taken:
The IMB recommends that:
Implanting surgeons / Implanting centres
• Identify women who were implanted with PIP silicone gel filled implants after 01 January 2001.
• Reassure them that there is no current evidence of health risk associated with the implants.
• Advise them that further information about the testing is available on the IMB, Therapeutic Goods Administration in Australia, Medicines And Healthcare products Regulatory Agency in UK and French Regulatory Agency AFSSAPs websites. (web addresses provided above).
GPs
Advise patients who are concerned about their PIP implants to consult their implanting surgeon/ implanting centres.
Further Information:
All adverse incidents relating to a medical device should be reported to the:
Health Products Regulatory Authority
Kevin O’Malley House
Earlsfort Centre
Earlsfort Terrace
Dublin 2
Telephone: +353-1-6764971
Fax: +353-1-6344033
E-mail:
vigilance@HPRA.ie
Website:
www.hpra.ie
Please Click here to download PDF version of Safety Notice
« Back
Date Printed: 04/10/2024