Tandem t:slim X2 Insulin Pump Priority 2 - Warning
16/12/2025
Medical devices
HPRA safety notice
| Point of information | Details |
| Device Manufacturer/Supplier Information | Tandem Diabetes Care/Air Liquide Healthcare Ireland |
| Target Audience | Patients / Users of these devices / Healthcare professionals and healthcare settings involved in the use and/or provision of these devices. |
| What devices are affected? | t:slim X2 insulin pump |
Problem or issue?
Tandem has identified that certain t:slim X2 insulin pumps include a speaker version that has a higher rate of malfunction which may appear as a 'Malfunction 16’ event on your pump. This is due to a wiring issue within the speaker. If a ‘Malfunction 16’ is declared on your pump, insulin delivery will stop, and the pump is no longer operational.
What is the risk?
If this issue occurs it could result in hyperglycemia due to the malfunction, resulting in the stoppage of insulin delivery as well as the real time continuous glucose monitoring, estimated glucose values, and blood glucose trend information. In severe cases of hyperglycemia, the user may require hospitalization or intervention from a medical professional.
Why is the HPRA issuing the safety notice?
The HPRA is issuing this safety notice to ensure adequate awareness of this issue and to ensure that users undertake the actions outlined in the manufacturers FSNs. The manufacturer will be releasing a software update in the coming months aimed at enhancing early detection of speaker failure.
What advice has the HPRA for users of these devices?
- Refer to the accompanying FSN’s and follow the instructions provided by the manufacturer. Please note, there are two FSN’s related to this issue; an FSN specific to pumps that have been in use less than 2 years or for users who are less than 18 years of age and an FSN for all other t: slim X2 insulin pump users.
- If you receive a ‘Malfunction 16’ alert on your insulin pump: a. Promptly acknowledge it by pressing “Silence Alarm”. b. Contact Air Liquide to report the malfunction and to initiate the return and replacement of your pump. c. Switch to your backup method of insulin delivery as directed by your health care professional.
- If you haven’t had a ‘Malfunction 16’ occur, you may continue using your Tandem pump as described in the User Guide, but with added precautions as the ‘Malfunction 16’ can occur at any time.
- Regularly check your blood sugar to ensure you are not having unexpectedly high or low readings.
- Be prepared with a backup method of insulin delivery as directed by your physician.
- If you are using a Dexcom sensor, use the Dexcom mobile app as a secondary means of CGM alerts and alarms.
- Report any safety issues associated with these devices to the manufacturer and to the HPRA.
Contact Information
| Manufacturer | Authorised Representative | Distributor | |
| Name: | Tandem Diabetes Care, Inc. | MDSS GmbH | Air Liquide Healthcare Ireland Ltd |
| Address: | 12400 High Bluff Drive, San Diego, CA,92130 | Schiffgraben 41, Hannover, Germany ,30175 | 25/28 N Wall Quay, North Wall, Dublin, D01 H104, Ireland |
| Email: | MTrier@tandemdiabetes.com | vigilance@mdssar.com | healthie@airliquide.ie |
| Telephone: | +1 805 297 9106 | +49 511 6262 8630 | 1800 12 4912 |
| Website: | www.tandemdiabetes.com | www.mdss.com | www.makingdiabeteseasier.com/uk |
HPRA Contact Information
All safety issues relating to medical devices should be reported to the HPRA using any of the following reporting options:
- HPRA online report form
- Telephone: +353-1-6764971
Further information on the role of the HPRA in relation to medical devices is available on the regulation of medical devices section of the HPRA website.
Any enquiries regarding this safety notice or any medical device safety issue should be emailed to us using our online contact form.
Documents
Opens in new window Tandem t:slim X2 Insulin Pump – HPRA Safety Notice – 16 December 2025 PDF : 203KB | 16/12/2025 Opens in new window Tandem t:slim X2 Insulin Pump – Field Safety Notice 1 – 25 August 2025 PDF : 173KB | 15/12/2025 Opens in new window Tandem t:slim X2 Insulin Pump – Field Safety Notice 2 – 13 November 2025 PDF : 164KB | 15/12/2025Product images