Zoll Powerheart G5 AEDs/Error code (EC) 501 - Priority 2 - Warning
Point of information | Details |
Device manufacturer/supplier information | Zoll Medical Corporation |
Target audience | Owners / users of these devices |
What devices are affected? | Powerheart G5 semi-automatic and automatic AEDs |
Problem or issue?
The HPRA is issuing this Safety Notice to raise awareness of an issue affecting the Powerheart G5 AED, whereby the device may not deliver therapy when needed.
After performing normally scheduled self-tests, Powerheart G5 AEDs may display Error Code (EC) 501 and be unable to deliver lifesaving therapy while in this state. The issue can be detected during routine inspection.
What is the risk?
If the error code outlined above is not detected, the device will be unable to deliver lifesaving therapy.
Why is the HPRA issuing this safety notice?
The manufacturer, via their local distributor Fleming Medical, has issued a field safety notice (FSN) in relation to this issue however the rate of acknowledgment from users in Ireland to date has been low. The HPRA is issuing this safety notice to raise awareness of the issue and ensure users undertake the actions outlined in the manufacturer’s FSN.
What advice has the HPRA for users of these devices?
- Refer to the accompanying FSN (linked at the bottom of the page) and follow instructions provided by the manufacturer.
- Ensure your organisation has a system in place for regular monitoring of these devices.
- If an error code EC 501 is detected, Zoll advises that the device should be removed from use.
- Appropriate action should be taken to resolve the issue, as outlined in the accompanying FSN, and replace the device before it is needed in an emergency situation.
- Store devices within the recommended humidity range.
- Report any safety issues associated with these devices to the manufacturer and the HPRA.
Contact information
Manufacturer | Distributor | |
Name | Zoll Medical Corporation | Fleming Medical Limited |
Address | 269 Mill Road, Chelmsford, MA 01824, USA | Corcanree Business Park, Dock Road, Limerick V94 K6W4 |
techservice-uk@zoll.com | info@flemingmedical.ie | |
Telephone | 0808 168 5000 or 0044 1928 241701 | 1800 307777 |
Website | www.zoll.com | www.flemingmedical.ie |
HPRA contact information
All safety issues relating to medical devices should be reported to the HPRA using any of the following reporting options:
- Telephone: +353-1-6764971
Documents
Opens in new window Powerheart G5 AEDs_Zoll_FSN for users PDF : 209KB | 05/06/2025 Opens in new window Powerheart G5 AEDs_Priority2_Warning_HPRA Safety Notice PDF : 112KB | 09/06/2025Product images